Abstract
We describe the design and performance of a beam line installed recently at Brookhaven intended primarily for high resolution photoemission studies on solid surfaces int he photon energy range 8-250 eV. The monochromator is a 6 m/160° toroidal grating instrument having three gratings 288, 822 and 2400 g/mm designed for a resolution as good as 0.2 eV over almost its entire range and considerably better than this at low energies. The photoelectron spectrometer is based on a hemispherical analyzer with a specially designed zoom lens, multianode detection, two rotational degrees of freedom and is capable of an ultimate resolution of 20 meV in energy and {precedes above single-line equals sign}0.01 A ̊-1 in momentum.
Original language | English (US) |
---|---|
Pages (from-to) | 85-90 |
Number of pages | 6 |
Journal | Nuclear Instruments and Methods In Physics Research |
Volume | 222 |
Issue number | 1-2 |
DOIs | |
State | Published - May 15 1984 |
Externally published | Yes |
ASJC Scopus subject areas
- General Engineering