Wireless Wearable Ultrasound Sensor on a Paper Substrate to Characterize Respiratory Behavior

Ang Chen, Andrew Joshua Halton, Rachel Diane Rhoades, Jayden Charles Booth, Xinhao Shi, Xiangli Bu, Ning Wu, Junseok Chae

Research output: Contribution to journalArticle

Abstract

Respiratory behavior contains crucial parameters to feature lung functionality, including respiratory rate, profile, and volume. The current well-adopted method to characterize respiratory behavior is spirometry using a spirometer, which is bulky, heavy, expensive, requires a trained provider to operate, and is incapable of continuous monitoring of respiratory behavior, which is often critical to assess chronic respiratory diseases. This work presents a wireless wearable sensor on a paper substrate that is capable of continuous monitoring of respiratory behavior and delivering the clinically relevant respiratory information to a smartphone. The wireless wearable sensor was attached on the midway of the xiphoid process and the costal margin, corresponding to the abdomen-apposed rib cage, based on the anatomical and experimental analysis. The sensor, with a footprint of 40 × 35 × 6 mm 3 and weighing 6.5 g, including a 2.7 g battery, consists of three subsystems, (i) ultrasound emitter, (ii) ultrasound receiver, and (iii) data acquisition and wireless transmitter. The sensor converts the linear strain at the wearing site to the lung volume change by measuring the change in ultrasound pressure as a function of the distance between the emitter and the receiver. The temporal lung volume change data, directly converted from the ultrasound pressure, is wirelessly transmitted to a smartphone where a custom-designed app computes to show volume-time and flow rate-volume loop graphs, standard respiratory analysis plots. The app analyzes the plots to show the clinically relevant respiratory behavioral parameters, such as forced vital capacity (FVC) and forced expiratory volume delivered in the first second (FEV 1 ). Potential user-induced error on sensor placement and temperature sensitivity were studied to demonstrate the sensor maintains its performance within a reasonable range of those variables. Eight volunteers were recruited to evaluate the sensor, which showed the mean deviation of the FEV 1 /FVC ratio in the range of 0.00-4.25% when benchmarked by the spirometer. The continuous measurement of respiratory behavioral parameters helps track the progression of the respiratory diseases, including asthma progression to provide alerts to relevant caregivers to seek needed timely treatment.

Original languageEnglish (US)
JournalACS Sensors
DOIs
StatePublished - Jan 1 2019

Fingerprint

Ultrasonics
sensors
Sensors
Substrates
spirometers
Pulmonary diseases
respiratory diseases
Smartphones
lungs
Application programs
progressions
emitters
Monitoring
respiratory rate
receivers
plots
Weighing
asthma
abdomen
Transmitters

Keywords

  • asthma control
  • paper substrate
  • ultrasound
  • wearable
  • wireless

ASJC Scopus subject areas

  • Bioengineering
  • Instrumentation
  • Process Chemistry and Technology
  • Fluid Flow and Transfer Processes

Cite this

Wireless Wearable Ultrasound Sensor on a Paper Substrate to Characterize Respiratory Behavior. / Chen, Ang; Halton, Andrew Joshua; Rhoades, Rachel Diane; Booth, Jayden Charles; Shi, Xinhao; Bu, Xiangli; Wu, Ning; Chae, Junseok.

In: ACS Sensors, 01.01.2019.

Research output: Contribution to journalArticle

Chen, Ang ; Halton, Andrew Joshua ; Rhoades, Rachel Diane ; Booth, Jayden Charles ; Shi, Xinhao ; Bu, Xiangli ; Wu, Ning ; Chae, Junseok. / Wireless Wearable Ultrasound Sensor on a Paper Substrate to Characterize Respiratory Behavior. In: ACS Sensors. 2019.
@article{82c4007dee9043489ee6538f939baa8e,
title = "Wireless Wearable Ultrasound Sensor on a Paper Substrate to Characterize Respiratory Behavior",
abstract = "Respiratory behavior contains crucial parameters to feature lung functionality, including respiratory rate, profile, and volume. The current well-adopted method to characterize respiratory behavior is spirometry using a spirometer, which is bulky, heavy, expensive, requires a trained provider to operate, and is incapable of continuous monitoring of respiratory behavior, which is often critical to assess chronic respiratory diseases. This work presents a wireless wearable sensor on a paper substrate that is capable of continuous monitoring of respiratory behavior and delivering the clinically relevant respiratory information to a smartphone. The wireless wearable sensor was attached on the midway of the xiphoid process and the costal margin, corresponding to the abdomen-apposed rib cage, based on the anatomical and experimental analysis. The sensor, with a footprint of 40 × 35 × 6 mm 3 and weighing 6.5 g, including a 2.7 g battery, consists of three subsystems, (i) ultrasound emitter, (ii) ultrasound receiver, and (iii) data acquisition and wireless transmitter. The sensor converts the linear strain at the wearing site to the lung volume change by measuring the change in ultrasound pressure as a function of the distance between the emitter and the receiver. The temporal lung volume change data, directly converted from the ultrasound pressure, is wirelessly transmitted to a smartphone where a custom-designed app computes to show volume-time and flow rate-volume loop graphs, standard respiratory analysis plots. The app analyzes the plots to show the clinically relevant respiratory behavioral parameters, such as forced vital capacity (FVC) and forced expiratory volume delivered in the first second (FEV 1 ). Potential user-induced error on sensor placement and temperature sensitivity were studied to demonstrate the sensor maintains its performance within a reasonable range of those variables. Eight volunteers were recruited to evaluate the sensor, which showed the mean deviation of the FEV 1 /FVC ratio in the range of 0.00-4.25{\%} when benchmarked by the spirometer. The continuous measurement of respiratory behavioral parameters helps track the progression of the respiratory diseases, including asthma progression to provide alerts to relevant caregivers to seek needed timely treatment.",
keywords = "asthma control, paper substrate, ultrasound, wearable, wireless",
author = "Ang Chen and Halton, {Andrew Joshua} and Rhoades, {Rachel Diane} and Booth, {Jayden Charles} and Xinhao Shi and Xiangli Bu and Ning Wu and Junseok Chae",
year = "2019",
month = "1",
day = "1",
doi = "10.1021/acssensors.9b00043",
language = "English (US)",
journal = "ACS Sensors",
issn = "2379-3694",
publisher = "American Chemical Society",

}

TY - JOUR

T1 - Wireless Wearable Ultrasound Sensor on a Paper Substrate to Characterize Respiratory Behavior

AU - Chen, Ang

AU - Halton, Andrew Joshua

AU - Rhoades, Rachel Diane

AU - Booth, Jayden Charles

AU - Shi, Xinhao

AU - Bu, Xiangli

AU - Wu, Ning

AU - Chae, Junseok

PY - 2019/1/1

Y1 - 2019/1/1

N2 - Respiratory behavior contains crucial parameters to feature lung functionality, including respiratory rate, profile, and volume. The current well-adopted method to characterize respiratory behavior is spirometry using a spirometer, which is bulky, heavy, expensive, requires a trained provider to operate, and is incapable of continuous monitoring of respiratory behavior, which is often critical to assess chronic respiratory diseases. This work presents a wireless wearable sensor on a paper substrate that is capable of continuous monitoring of respiratory behavior and delivering the clinically relevant respiratory information to a smartphone. The wireless wearable sensor was attached on the midway of the xiphoid process and the costal margin, corresponding to the abdomen-apposed rib cage, based on the anatomical and experimental analysis. The sensor, with a footprint of 40 × 35 × 6 mm 3 and weighing 6.5 g, including a 2.7 g battery, consists of three subsystems, (i) ultrasound emitter, (ii) ultrasound receiver, and (iii) data acquisition and wireless transmitter. The sensor converts the linear strain at the wearing site to the lung volume change by measuring the change in ultrasound pressure as a function of the distance between the emitter and the receiver. The temporal lung volume change data, directly converted from the ultrasound pressure, is wirelessly transmitted to a smartphone where a custom-designed app computes to show volume-time and flow rate-volume loop graphs, standard respiratory analysis plots. The app analyzes the plots to show the clinically relevant respiratory behavioral parameters, such as forced vital capacity (FVC) and forced expiratory volume delivered in the first second (FEV 1 ). Potential user-induced error on sensor placement and temperature sensitivity were studied to demonstrate the sensor maintains its performance within a reasonable range of those variables. Eight volunteers were recruited to evaluate the sensor, which showed the mean deviation of the FEV 1 /FVC ratio in the range of 0.00-4.25% when benchmarked by the spirometer. The continuous measurement of respiratory behavioral parameters helps track the progression of the respiratory diseases, including asthma progression to provide alerts to relevant caregivers to seek needed timely treatment.

AB - Respiratory behavior contains crucial parameters to feature lung functionality, including respiratory rate, profile, and volume. The current well-adopted method to characterize respiratory behavior is spirometry using a spirometer, which is bulky, heavy, expensive, requires a trained provider to operate, and is incapable of continuous monitoring of respiratory behavior, which is often critical to assess chronic respiratory diseases. This work presents a wireless wearable sensor on a paper substrate that is capable of continuous monitoring of respiratory behavior and delivering the clinically relevant respiratory information to a smartphone. The wireless wearable sensor was attached on the midway of the xiphoid process and the costal margin, corresponding to the abdomen-apposed rib cage, based on the anatomical and experimental analysis. The sensor, with a footprint of 40 × 35 × 6 mm 3 and weighing 6.5 g, including a 2.7 g battery, consists of three subsystems, (i) ultrasound emitter, (ii) ultrasound receiver, and (iii) data acquisition and wireless transmitter. The sensor converts the linear strain at the wearing site to the lung volume change by measuring the change in ultrasound pressure as a function of the distance between the emitter and the receiver. The temporal lung volume change data, directly converted from the ultrasound pressure, is wirelessly transmitted to a smartphone where a custom-designed app computes to show volume-time and flow rate-volume loop graphs, standard respiratory analysis plots. The app analyzes the plots to show the clinically relevant respiratory behavioral parameters, such as forced vital capacity (FVC) and forced expiratory volume delivered in the first second (FEV 1 ). Potential user-induced error on sensor placement and temperature sensitivity were studied to demonstrate the sensor maintains its performance within a reasonable range of those variables. Eight volunteers were recruited to evaluate the sensor, which showed the mean deviation of the FEV 1 /FVC ratio in the range of 0.00-4.25% when benchmarked by the spirometer. The continuous measurement of respiratory behavioral parameters helps track the progression of the respiratory diseases, including asthma progression to provide alerts to relevant caregivers to seek needed timely treatment.

KW - asthma control

KW - paper substrate

KW - ultrasound

KW - wearable

KW - wireless

UR - http://www.scopus.com/inward/record.url?scp=85063490592&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85063490592&partnerID=8YFLogxK

U2 - 10.1021/acssensors.9b00043

DO - 10.1021/acssensors.9b00043

M3 - Article

C2 - 30855133

AN - SCOPUS:85063490592

JO - ACS Sensors

JF - ACS Sensors

SN - 2379-3694

ER -