T1 - Oxygen isotope measurements of seawater ( H218O/H216O)
T2 - A comparison of cavity ring-down spectroscopy (CRDS) and isotope ratio mass spectrometry (IRMS)
AU - Walker, S. A.
AU - Azetsu-Scott, K.
AU - Normandeau, C.
AU - Kelley, D. E.
AU - Friedrich, R.
AU - Newton, R.
AU - Schlosser, P.
AU - Mckay, J. L.
AU - Abdi, W.
AU - Kerrigan, E.
AU - Craig, S. E.
AU - Wallace, D. W.R.
N1 - Publisher Copyright:
© 2016 Association for the Sciences of Limnology and Oceanography.
PY - 2016/1/1
Y1 - 2016/1/1
N2 - Laser-based spectroscopic techniques, such as cavity ring-down spectroscopy (CRDS), provide a new, cost effective and more widely available approach to measure the oxygen isotope ratio in water molecules, H218O/ H216O (δ18O), and are used increasingly to measure δ18O in the world's oceans. Here, we present results from an interlaboratory comparison designed to evaluate the quality of CRDS-derived measurements, and their consistency with values measured by isotope ratio mass spectrometry (IRMS). We also discuss the influence of salt on instrument performance and sample throughput for the analysis of seawater samples. This study compared measurements of δ18O from natural samples with a wide range of salinities (0, 29.4, and 34.6) performed by four independent labs: two using CRDS and two using IRMS. We also compared δ18O measurements of Northeast Atlantic Deep Water collected in 2013, 2012, 2009, and 1995 from the AR7W repeat hydrography transect across the Labrador Sea. The within-lab precision of ocean-based CRDS measurements is seen to approach 0.03‰, which is better than the manufacturer's typically stated analytical precision (around +/- 0.05‰), and comparable to that achievable with IRMS. The interlaboratory differences of measurements (highest-lowest) reported by the four labs is taken as an indicator of overall accuracy, and is estimated conservatively as being<0.1‰, with the potential to approach 0.05‰. Overall, these results show that CRDS based 18O measurements of seawater can be equivalent to high-quality measurements by IRMS.
AB - Laser-based spectroscopic techniques, such as cavity ring-down spectroscopy (CRDS), provide a new, cost effective and more widely available approach to measure the oxygen isotope ratio in water molecules, H218O/ H216O (δ18O), and are used increasingly to measure δ18O in the world's oceans. Here, we present results from an interlaboratory comparison designed to evaluate the quality of CRDS-derived measurements, and their consistency with values measured by isotope ratio mass spectrometry (IRMS). We also discuss the influence of salt on instrument performance and sample throughput for the analysis of seawater samples. This study compared measurements of δ18O from natural samples with a wide range of salinities (0, 29.4, and 34.6) performed by four independent labs: two using CRDS and two using IRMS. We also compared δ18O measurements of Northeast Atlantic Deep Water collected in 2013, 2012, 2009, and 1995 from the AR7W repeat hydrography transect across the Labrador Sea. The within-lab precision of ocean-based CRDS measurements is seen to approach 0.03‰, which is better than the manufacturer's typically stated analytical precision (around +/- 0.05‰), and comparable to that achievable with IRMS. The interlaboratory differences of measurements (highest-lowest) reported by the four labs is taken as an indicator of overall accuracy, and is estimated conservatively as being<0.1‰, with the potential to approach 0.05‰. Overall, these results show that CRDS based 18O measurements of seawater can be equivalent to high-quality measurements by IRMS.
UR - http://www.scopus.com/inward/record.url?scp=85000770402&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85000770402&partnerID=8YFLogxK
U2 - 10.1002/lom3.10067
DO - 10.1002/lom3.10067
M3 - Article
AN - SCOPUS:85000770402
SN - 1541-5856
VL - 14
SP - 31
EP - 38
JO - Limnology and Oceanography: Methods
JF - Limnology and Oceanography: Methods
IS - 1
ER -