MO‐E‐204B‐04: Interventional Fluoroscopy Procedure Skin Dose Assessment Using DICOM Dose Structured Reports

B. Schueler, W. Pavlicek, Y. Khodadadegan, M. Zhang, R. Paden, K. Fetterly, Teresa Wu, S. Langer

Research output: Contribution to journalArticle

Abstract

Purpose: Patient skin injury due to radiation exposure during interventional fluoroscopy procedures is a concern. The FDA has recommended dose be monitored, along with the dose distribution on the patient's skin. Modern fluoroscopy equipment provides a readout of the accumulated reference point air kerma. However, this parameter is a poor estimate of localized skin dose since it does not take into account beam movement. Acquisition information needed to calculate skin dose can be recorded as a DICOM Dose Structured Report (Dose SR), which has recently become available on new fluoroscopy equipment. We have developed a system that uses the Dose SR to produce a skin dose map, along with determination of the peak skin dose. Method and Materials: The Dose SR provides data for each irradiation event (fluoroscopy footswitch or acquisition): kerma‐area‐product, reference point air kerma, C‐arm angulation and table position. Using the position of the patient on the table, backscatter and table attenuation, the air kerma to 1‐cm2 regions on the patient's surface is calculated. The patient surface is approximated as a mathematical model using SolidWorks simulation software, with model size varied based on patient parameters. Results: The method was validated using a Siemens Axiom Artis angiography system. Measurements made with an ionization chamber and lucite phantom were compared to calculated skin dose values for a range of C‐arm angles, table positions and exposure levels. Errors ranged from 2–10% with higher errors corresponding to increased C‐arm angulation. Conclusion: A system has been developed to estimate skin dose and generate skin dose maps from an interventional procedure Dose SR. The system allows for a more precise determination of localized skin dose and assists with planning appropriate entry ports for future procedures to minimize skin injury. Application to any vendor platform providing a Dose SR is possible.

Original languageEnglish (US)
Pages (from-to)3359
Number of pages1
JournalMedical Physics
Volume37
Issue number6
DOIs
StatePublished - 2010
Externally publishedYes

Fingerprint

Fluoroscopy
Skin
Air
Equipment and Supplies
Wounds and Injuries
Polymethyl Methacrylate
Angiography
Theoretical Models
Software

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Schueler, B., Pavlicek, W., Khodadadegan, Y., Zhang, M., Paden, R., Fetterly, K., ... Langer, S. (2010). MO‐E‐204B‐04: Interventional Fluoroscopy Procedure Skin Dose Assessment Using DICOM Dose Structured Reports. Medical Physics, 37(6), 3359. https://doi.org/10.1118/1.3469130

MO‐E‐204B‐04 : Interventional Fluoroscopy Procedure Skin Dose Assessment Using DICOM Dose Structured Reports. / Schueler, B.; Pavlicek, W.; Khodadadegan, Y.; Zhang, M.; Paden, R.; Fetterly, K.; Wu, Teresa; Langer, S.

In: Medical Physics, Vol. 37, No. 6, 2010, p. 3359.

Research output: Contribution to journalArticle

Schueler, B, Pavlicek, W, Khodadadegan, Y, Zhang, M, Paden, R, Fetterly, K, Wu, T & Langer, S 2010, 'MO‐E‐204B‐04: Interventional Fluoroscopy Procedure Skin Dose Assessment Using DICOM Dose Structured Reports', Medical Physics, vol. 37, no. 6, pp. 3359. https://doi.org/10.1118/1.3469130
Schueler, B. ; Pavlicek, W. ; Khodadadegan, Y. ; Zhang, M. ; Paden, R. ; Fetterly, K. ; Wu, Teresa ; Langer, S. / MO‐E‐204B‐04 : Interventional Fluoroscopy Procedure Skin Dose Assessment Using DICOM Dose Structured Reports. In: Medical Physics. 2010 ; Vol. 37, No. 6. pp. 3359.
@article{af4dd2999f464860aac35e47dfbae792,
title = "MO‐E‐204B‐04: Interventional Fluoroscopy Procedure Skin Dose Assessment Using DICOM Dose Structured Reports",
abstract = "Purpose: Patient skin injury due to radiation exposure during interventional fluoroscopy procedures is a concern. The FDA has recommended dose be monitored, along with the dose distribution on the patient's skin. Modern fluoroscopy equipment provides a readout of the accumulated reference point air kerma. However, this parameter is a poor estimate of localized skin dose since it does not take into account beam movement. Acquisition information needed to calculate skin dose can be recorded as a DICOM Dose Structured Report (Dose SR), which has recently become available on new fluoroscopy equipment. We have developed a system that uses the Dose SR to produce a skin dose map, along with determination of the peak skin dose. Method and Materials: The Dose SR provides data for each irradiation event (fluoroscopy footswitch or acquisition): kerma‐area‐product, reference point air kerma, C‐arm angulation and table position. Using the position of the patient on the table, backscatter and table attenuation, the air kerma to 1‐cm2 regions on the patient's surface is calculated. The patient surface is approximated as a mathematical model using SolidWorks simulation software, with model size varied based on patient parameters. Results: The method was validated using a Siemens Axiom Artis angiography system. Measurements made with an ionization chamber and lucite phantom were compared to calculated skin dose values for a range of C‐arm angles, table positions and exposure levels. Errors ranged from 2–10{\%} with higher errors corresponding to increased C‐arm angulation. Conclusion: A system has been developed to estimate skin dose and generate skin dose maps from an interventional procedure Dose SR. The system allows for a more precise determination of localized skin dose and assists with planning appropriate entry ports for future procedures to minimize skin injury. Application to any vendor platform providing a Dose SR is possible.",
author = "B. Schueler and W. Pavlicek and Y. Khodadadegan and M. Zhang and R. Paden and K. Fetterly and Teresa Wu and S. Langer",
year = "2010",
doi = "10.1118/1.3469130",
language = "English (US)",
volume = "37",
pages = "3359",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - MO‐E‐204B‐04

T2 - Interventional Fluoroscopy Procedure Skin Dose Assessment Using DICOM Dose Structured Reports

AU - Schueler, B.

AU - Pavlicek, W.

AU - Khodadadegan, Y.

AU - Zhang, M.

AU - Paden, R.

AU - Fetterly, K.

AU - Wu, Teresa

AU - Langer, S.

PY - 2010

Y1 - 2010

N2 - Purpose: Patient skin injury due to radiation exposure during interventional fluoroscopy procedures is a concern. The FDA has recommended dose be monitored, along with the dose distribution on the patient's skin. Modern fluoroscopy equipment provides a readout of the accumulated reference point air kerma. However, this parameter is a poor estimate of localized skin dose since it does not take into account beam movement. Acquisition information needed to calculate skin dose can be recorded as a DICOM Dose Structured Report (Dose SR), which has recently become available on new fluoroscopy equipment. We have developed a system that uses the Dose SR to produce a skin dose map, along with determination of the peak skin dose. Method and Materials: The Dose SR provides data for each irradiation event (fluoroscopy footswitch or acquisition): kerma‐area‐product, reference point air kerma, C‐arm angulation and table position. Using the position of the patient on the table, backscatter and table attenuation, the air kerma to 1‐cm2 regions on the patient's surface is calculated. The patient surface is approximated as a mathematical model using SolidWorks simulation software, with model size varied based on patient parameters. Results: The method was validated using a Siemens Axiom Artis angiography system. Measurements made with an ionization chamber and lucite phantom were compared to calculated skin dose values for a range of C‐arm angles, table positions and exposure levels. Errors ranged from 2–10% with higher errors corresponding to increased C‐arm angulation. Conclusion: A system has been developed to estimate skin dose and generate skin dose maps from an interventional procedure Dose SR. The system allows for a more precise determination of localized skin dose and assists with planning appropriate entry ports for future procedures to minimize skin injury. Application to any vendor platform providing a Dose SR is possible.

AB - Purpose: Patient skin injury due to radiation exposure during interventional fluoroscopy procedures is a concern. The FDA has recommended dose be monitored, along with the dose distribution on the patient's skin. Modern fluoroscopy equipment provides a readout of the accumulated reference point air kerma. However, this parameter is a poor estimate of localized skin dose since it does not take into account beam movement. Acquisition information needed to calculate skin dose can be recorded as a DICOM Dose Structured Report (Dose SR), which has recently become available on new fluoroscopy equipment. We have developed a system that uses the Dose SR to produce a skin dose map, along with determination of the peak skin dose. Method and Materials: The Dose SR provides data for each irradiation event (fluoroscopy footswitch or acquisition): kerma‐area‐product, reference point air kerma, C‐arm angulation and table position. Using the position of the patient on the table, backscatter and table attenuation, the air kerma to 1‐cm2 regions on the patient's surface is calculated. The patient surface is approximated as a mathematical model using SolidWorks simulation software, with model size varied based on patient parameters. Results: The method was validated using a Siemens Axiom Artis angiography system. Measurements made with an ionization chamber and lucite phantom were compared to calculated skin dose values for a range of C‐arm angles, table positions and exposure levels. Errors ranged from 2–10% with higher errors corresponding to increased C‐arm angulation. Conclusion: A system has been developed to estimate skin dose and generate skin dose maps from an interventional procedure Dose SR. The system allows for a more precise determination of localized skin dose and assists with planning appropriate entry ports for future procedures to minimize skin injury. Application to any vendor platform providing a Dose SR is possible.

UR - http://www.scopus.com/inward/record.url?scp=85024777172&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024777172&partnerID=8YFLogxK

U2 - 10.1118/1.3469130

DO - 10.1118/1.3469130

M3 - Article

AN - SCOPUS:85024777172

VL - 37

SP - 3359

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -