TY - JOUR
T1 - Interaural level differences and sound source localization for bilateral cochlear implant patients
AU - Dorman, Michael
AU - Loiselle, Louise
AU - Stohl, Josh
AU - Yost, William
AU - Spahr, Anthony
AU - Brown, Chris
AU - Cook, Sarah
N1 - Publisher Copyright:
Copyright © 2014 by Lippincott Williams & Wilkins · Printed in the U.S.A.
PY - 2014
Y1 - 2014
N2 - Objectives: The aims of this study were (i) to determine the magnitude of the interaural level differences (ILDs) that remain after cochlear implant (CI) signal processing and (ii) to relate the ILDs to the pattern of errors for sound source localization on the horizontal plane. Design: The listeners were 16 bilateral CI patients fitted with MED-EL CIs and 34 normal-hearing listeners. The stimuli were wideband, high-pass, and low-pass noise signals. ILDs were calculated by passing signals, filtered by head-related transfer functions (HRTFs) to a Matlab simulation of MED-EL signal processing. Results: For the wideband signal and high-pass signals, maximum ILDs of 15 to 17 dB in the input signal were reduced to 3 to 4 dB after CI signal processing. For the low-pass signal, ILDs were reduced to 1 to 2 dB. For wideband and high-pass signals, the largest ILDs for ±15 degree speaker locations were between 0.4 and 0.7 dB; for the ±30 degree speaker locations between 0.9 and 1.3 dB; for the 45 degree speaker locations between 2.4 and 2.9 dB; for the ±60 degree speaker locations, between 3.2 and 4.1 dB; and for the ±75 degree speaker locations between 2.7 and 3.4 dB. All of the CI patients in all the stimulus conditions showed poorer localization than the normal-hearing listeners. Localization accuracy for the CI patients was best for the wideband and high-pass signals and was poorest for the low-pass signal. Conclusions: Localization accuracy was related to the magnitude of the ILD cues available to the normal-hearing listeners and CI patients. The pattern of localization errors for the CI patients was related to the magnitude of the ILD differences among loudspeaker locations. The error patterns for the wideband and high-pass signals, suggest that, for the conditions of this experiment, patients, on an average, sorted signals on the horizontal plane into four sectors-on each side of the midline, one sector including 0, 15, and possibly 30 degree speaker locations, and a sector from 45 degree speaker locations to 75 degree speaker locations. The resolution within a sector was relatively poor.
AB - Objectives: The aims of this study were (i) to determine the magnitude of the interaural level differences (ILDs) that remain after cochlear implant (CI) signal processing and (ii) to relate the ILDs to the pattern of errors for sound source localization on the horizontal plane. Design: The listeners were 16 bilateral CI patients fitted with MED-EL CIs and 34 normal-hearing listeners. The stimuli were wideband, high-pass, and low-pass noise signals. ILDs were calculated by passing signals, filtered by head-related transfer functions (HRTFs) to a Matlab simulation of MED-EL signal processing. Results: For the wideband signal and high-pass signals, maximum ILDs of 15 to 17 dB in the input signal were reduced to 3 to 4 dB after CI signal processing. For the low-pass signal, ILDs were reduced to 1 to 2 dB. For wideband and high-pass signals, the largest ILDs for ±15 degree speaker locations were between 0.4 and 0.7 dB; for the ±30 degree speaker locations between 0.9 and 1.3 dB; for the 45 degree speaker locations between 2.4 and 2.9 dB; for the ±60 degree speaker locations, between 3.2 and 4.1 dB; and for the ±75 degree speaker locations between 2.7 and 3.4 dB. All of the CI patients in all the stimulus conditions showed poorer localization than the normal-hearing listeners. Localization accuracy for the CI patients was best for the wideband and high-pass signals and was poorest for the low-pass signal. Conclusions: Localization accuracy was related to the magnitude of the ILD cues available to the normal-hearing listeners and CI patients. The pattern of localization errors for the CI patients was related to the magnitude of the ILD differences among loudspeaker locations. The error patterns for the wideband and high-pass signals, suggest that, for the conditions of this experiment, patients, on an average, sorted signals on the horizontal plane into four sectors-on each side of the midline, one sector including 0, 15, and possibly 30 degree speaker locations, and a sector from 45 degree speaker locations to 75 degree speaker locations. The resolution within a sector was relatively poor.
KW - Cochlear implants
KW - ILD
KW - Sound source localization
UR - http://www.scopus.com/inward/record.url?scp=84927642643&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84927642643&partnerID=8YFLogxK
U2 - 10.1097/AUD.0000000000000057
DO - 10.1097/AUD.0000000000000057
M3 - Article
C2 - 25127322
AN - SCOPUS:84927642643
SN - 0196-0202
VL - 35
SP - 633
EP - 640
JO - Ear and hearing
JF - Ear and hearing
IS - 6
ER -