In vivo quantification of accumulating abdominal fluid using an electrical impedance tomography hemiarray

Aaron S. Tucker, Edward A. Ross, Jennifer Paugh-Miller, Rosalind J. Sadleir

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

A new method to image and quantify intra-abdominal haemorrhage using electrical impedance tomography (EIT) was tested in vivo. Supine peritoneal dialysis patients were monitored using an 8-electrode hemiarray placed on the anterior abdomen. EIT measurements were recorded using the EPack II data acquisition system before, during, and after the administration of dialysate. The amount of dialysate infused was recorded synchronous with EIT measurements and used as a control. Tomographic images of impedance change were reconstructed using a weighted, sensitivity-based method and then post-processed to obtain a quantitative estimate of the total dialysate volume added and the rate of dialysate administration. Our preliminary study included two subjects, one male and one female, each of whom participated for two sessions spaced about 6 months apart. Data collected from these sessions indicated that with an in vivo SNR of about 35 dB the EPack II can detect accumulations larger than about 100 ml, with a quantification uncertainty of about 50 ml. The rate of accumulation was determined in less than 2 min. This method shows promise for automated detection of other pathologies, eg ascites, and is adaptable to detecting conductive accumulations in other anatomy.

Original languageEnglish (US)
Pages (from-to)151-165
Number of pages15
JournalPhysiological Measurement
Volume32
Issue number2
DOIs
StatePublished - Feb 2011
Externally publishedYes

Keywords

  • Abdominal haemorrhage
  • Electrical impedance tomography
  • Electrode array
  • Flow rate
  • Noninvasive monitoring
  • Peritoneal dialysis

ASJC Scopus subject areas

  • Biophysics
  • Physiology
  • Biomedical Engineering
  • Physiology (medical)

Fingerprint

Dive into the research topics of 'In vivo quantification of accumulating abdominal fluid using an electrical impedance tomography hemiarray'. Together they form a unique fingerprint.

Cite this