T1 - Direct comparison of two actigraphy devices with polysomnographically recorded naps in healthy young adults
AU - Cellini, Nicola
AU - Buman, Matthew
AU - Mcdevitt, Elizabeth A.
AU - Ricker, Ashley A.
AU - Mednick, Sara C.
PY - 2013/6
Y1 - 2013/6
N2 - The last 20 yrs have seen a marked increase in studies utilizing actigraphy in free-living environments. The aim of the present study is to directly compare two commercially available actigraph devices with concurrent polysomnography (PSG) during a daytime nap in healthy young adults. Thirty healthy young adults, ages 18-31 (mean 20.77 yrs, SD 3.14 yrs) simultaneously wore AW-64 and GT3X+ devices during a polysomnographically recorded nap. Mann-Whitney U (M-U) test, intraclass correlation coefficients, and Bland-Altman statistic were used to compare total sleep time (TST), sleep onset latency (SOL), wake after sleep onset (WASO), and sleep efficiency (SE) between the two actigraphs and PSG. Epoch-by-epoch (EBE) agreement was calculated to determine accuracy, sensitivity, specificity, predictive values for sleep (PVS) and wake (PVW), and kappa and prevalence- and bias-adjusted kappa (PABAK) coefficients. All frequency settings provided by the devices were examined. For both actigraphs, EBE analysis found accuracy, sensitivity, specificity, PVS, and PVW comparable to previous reports of other similar devices. Kappa and PABAK coefficients showed moderate to high agreement with PSG depending on device settings. The GT3X+ overestimated TST and SE, and underestimated SOL and WASO, whereas no significant difference was found between AW-64 and PSG. However, GT3X+ showed overall better EBE agreements to PSG than AW-64. We conclude that both actigraphs are valid and reliable devices for detecting sleep/wake diurnal patterns. The choice between devices should be based on several parameters as reliability, cost of the device, scoring algorithm, target population, experimental condition, and aims of the study (e.g., sleep and/or physical activity).
AB - The last 20 yrs have seen a marked increase in studies utilizing actigraphy in free-living environments. The aim of the present study is to directly compare two commercially available actigraph devices with concurrent polysomnography (PSG) during a daytime nap in healthy young adults. Thirty healthy young adults, ages 18-31 (mean 20.77 yrs, SD 3.14 yrs) simultaneously wore AW-64 and GT3X+ devices during a polysomnographically recorded nap. Mann-Whitney U (M-U) test, intraclass correlation coefficients, and Bland-Altman statistic were used to compare total sleep time (TST), sleep onset latency (SOL), wake after sleep onset (WASO), and sleep efficiency (SE) between the two actigraphs and PSG. Epoch-by-epoch (EBE) agreement was calculated to determine accuracy, sensitivity, specificity, predictive values for sleep (PVS) and wake (PVW), and kappa and prevalence- and bias-adjusted kappa (PABAK) coefficients. All frequency settings provided by the devices were examined. For both actigraphs, EBE analysis found accuracy, sensitivity, specificity, PVS, and PVW comparable to previous reports of other similar devices. Kappa and PABAK coefficients showed moderate to high agreement with PSG depending on device settings. The GT3X+ overestimated TST and SE, and underestimated SOL and WASO, whereas no significant difference was found between AW-64 and PSG. However, GT3X+ showed overall better EBE agreements to PSG than AW-64. We conclude that both actigraphs are valid and reliable devices for detecting sleep/wake diurnal patterns. The choice between devices should be based on several parameters as reliability, cost of the device, scoring algorithm, target population, experimental condition, and aims of the study (e.g., sleep and/or physical activity).
KW - Actigraphy
KW - Bland-Altman
KW - Nap
KW - Polysomnography
KW - Validation
UR - http://www.scopus.com/inward/record.url?scp=84879363026&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84879363026&partnerID=8YFLogxK
U2 - 10.3109/07420528.2013.782312
DO - 10.3109/07420528.2013.782312
M3 - Article
C2 - 23721120
AN - SCOPUS:84879363026
SN - 0743-9539
VL - 30
SP - 691
EP - 698
JO - Annual Review of Chronopharmacology
JF - Annual Review of Chronopharmacology
IS - 5
ER -