T1 - Comparative study of a novel portable indirect calorimeter to a reference breath-by-breath instrument and its use in telemedicine settings
AU - Mora, S. Jimena
AU - Sprowls, Mark
AU - Tipparaju, Vishal V.
AU - Wheatley-Guy, Courtney M.
AU - Kulick, Doina
AU - Johnson, Bruce
AU - Xiaojun, Xian
AU - Forzani, Erica
N1 - Funding Information:
The authors acknowledge the participants for compliance to protocols. In addition, we acknowledge NIH R03 funding from NIBIB ( EB027336-02 ) and, A. J. and . Sigismunda Palumbo Charitable Trust .
Publisher Copyright:
© 2021 European Society for Clinical Nutrition and Metabolism
PY - 2021/12
Y1 - 2021/12
N2 - Background & aims: Resting Energy Expenditure (REE) quantitatively describes the calories used to support body function (e.g. breathing, blood circulation, etc.) at resting condition. Assessment of the REE is essential for successful weight management and the understanding of metabolic health. REE is typically determined via indirect calorimetry. Current biomedical indirect calorimetry technologies, utilizing assessment of oxygen consumption (VO2) and carbon dioxide production (VCO2) rates (which are typically in the form factor of a metabolic cart) are bulky and require on-site calibration and/or trained professionals to operate. We introduce a novel wearable medical device with FDA clearance to determine REE accurately, portable, and user-friendly format, which can be used both by health professionals in a clinical environment and by the patient at home. Previously, we have reported the validation of Breezing Med (also named as Breezing Pro™) through Douglas Bag Method, a gold standard for gas exchange measurement, and excellent agreement has been found between the two methods for the determination of REE, VO2, and VCO2 rates (Mora et al., 2020). Now we present the validation of Breezing Med against Medical Graphics (MGC) CPX Ultima™, a FDA 510 k cleared metabolic cart, which principle is based on breath-by-breath analysis. In addition, we present Breezing Med as a tool for daily measurement of metabolic rate by the lay person at home. Methods: A) The validation study was executed via parallel measurement of 20 healthy participants under resting conditions using both the Breezing Med and the MGC Ultima CPX™ (10 min test). B) Breezing Med measurements were carried out by six subjects at home during stay-at-home order due to COVID-19 for 30 days. Results: A) The resulting measurements from both devices was compared with correlation slope's and R-squared coefficients close to 1. B) Results were recorded and analyzed for variability. The pilot study demonstrated the advantage of Breezing Med device to be easy-to-use at home by lay people, which make the valuable device for telemedicine applications related to weight management from home. Conclusions: This result shows that the MGC Ultima CPX™ and Breezing Med are substantially equivalent for REE measurement; and an advantage of this device for metabolic assessment under the current COVID-19 pandemic situation, for people with impaired physical mobility, and for those who lives in rural areas or face impediments that limit physical access to care.
AB - Background & aims: Resting Energy Expenditure (REE) quantitatively describes the calories used to support body function (e.g. breathing, blood circulation, etc.) at resting condition. Assessment of the REE is essential for successful weight management and the understanding of metabolic health. REE is typically determined via indirect calorimetry. Current biomedical indirect calorimetry technologies, utilizing assessment of oxygen consumption (VO2) and carbon dioxide production (VCO2) rates (which are typically in the form factor of a metabolic cart) are bulky and require on-site calibration and/or trained professionals to operate. We introduce a novel wearable medical device with FDA clearance to determine REE accurately, portable, and user-friendly format, which can be used both by health professionals in a clinical environment and by the patient at home. Previously, we have reported the validation of Breezing Med (also named as Breezing Pro™) through Douglas Bag Method, a gold standard for gas exchange measurement, and excellent agreement has been found between the two methods for the determination of REE, VO2, and VCO2 rates (Mora et al., 2020). Now we present the validation of Breezing Med against Medical Graphics (MGC) CPX Ultima™, a FDA 510 k cleared metabolic cart, which principle is based on breath-by-breath analysis. In addition, we present Breezing Med as a tool for daily measurement of metabolic rate by the lay person at home. Methods: A) The validation study was executed via parallel measurement of 20 healthy participants under resting conditions using both the Breezing Med and the MGC Ultima CPX™ (10 min test). B) Breezing Med measurements were carried out by six subjects at home during stay-at-home order due to COVID-19 for 30 days. Results: A) The resulting measurements from both devices was compared with correlation slope's and R-squared coefficients close to 1. B) Results were recorded and analyzed for variability. The pilot study demonstrated the advantage of Breezing Med device to be easy-to-use at home by lay people, which make the valuable device for telemedicine applications related to weight management from home. Conclusions: This result shows that the MGC Ultima CPX™ and Breezing Med are substantially equivalent for REE measurement; and an advantage of this device for metabolic assessment under the current COVID-19 pandemic situation, for people with impaired physical mobility, and for those who lives in rural areas or face impediments that limit physical access to care.
KW - Breezing Med
KW - Breezing Pro
KW - Metabolic tracker
KW - Point of care
KW - Resting energy expenditure
KW - Telemedicine
UR - http://www.scopus.com/inward/record.url?scp=85116693424&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85116693424&partnerID=8YFLogxK
U2 - 10.1016/j.clnesp.2021.09.731
DO - 10.1016/j.clnesp.2021.09.731
M3 - Article
C2 - 34857221
AN - SCOPUS:85116693424
SN - 2405-4577
VL - 46
SP - 361
EP - 366
JO - Clinical Nutrition ESPEN
JF - Clinical Nutrition ESPEN
ER -