TY - JOUR
T1 - A New Method for Simulating Embolic Coils as Heterogeneous Porous Media
AU - Yadollahi-Farsani, Hooman
AU - Herrmann, Marcus
AU - Frakes, David
AU - Chong, Brian
N1 - Funding Information:
We would like to acknowledge the support from the National Science Foundation (Award #1512553). We would like to thank Endovantage, LLC for providing us with the numerical coil deployments.
Publisher Copyright:
© 2018, Biomedical Engineering Society.
PY - 2019/3/15
Y1 - 2019/3/15
N2 - Purpose: To gain insight into the influence of coils on aneurysmal hemodynamics, computational fluid dynamics (CFD) can be used. Conventional methods of modeling coils consider the explicit geometry of the deployed devices within the aneurysm and discretize the fluid domain. However, the complex geometry of a coil mass leads to cumbersome domain discretization along with a significant number of mesh elements. These problems have motivated a homogeneous porous medium coil model, whereby the explicit geometry of the coils is greatly simplified, and relevant homogeneous porous medium parameters are approximated. Unfortunately, since the coils are not distributed uniformly in the aneurysm, the homogeneity assumption is no longer valid. Methods: In this paper, a novel heterogeneous porous medium approach is introduced. To verify the model, we performed CFD simulations to calculate the pressure drop caused by actual deployed coils in a straight cylinder. Next, we considered three different anatomical aneurysm geometries virtually treated with coils and studied the hemodynamics using the presented heterogeneous porous medium model. Results: We show that the blood kinetic energy predicted by the heterogeneous model is in strong agreement with the conventional approach. The homogeneity assumption, on the other hand, significantly over-predicts the blood kinetic energy within the aneurysmal sac. Conclusions: These results indicate that the benefits of the porous medium assumption can be retained if a heterogeneous approach is applied. Implementation of the presented method led to a substantial reduction in the total number of mesh elements compared to the conventional method, and greater accuracy was enabled by considering heterogeneity compared to the homogenous approach.
AB - Purpose: To gain insight into the influence of coils on aneurysmal hemodynamics, computational fluid dynamics (CFD) can be used. Conventional methods of modeling coils consider the explicit geometry of the deployed devices within the aneurysm and discretize the fluid domain. However, the complex geometry of a coil mass leads to cumbersome domain discretization along with a significant number of mesh elements. These problems have motivated a homogeneous porous medium coil model, whereby the explicit geometry of the coils is greatly simplified, and relevant homogeneous porous medium parameters are approximated. Unfortunately, since the coils are not distributed uniformly in the aneurysm, the homogeneity assumption is no longer valid. Methods: In this paper, a novel heterogeneous porous medium approach is introduced. To verify the model, we performed CFD simulations to calculate the pressure drop caused by actual deployed coils in a straight cylinder. Next, we considered three different anatomical aneurysm geometries virtually treated with coils and studied the hemodynamics using the presented heterogeneous porous medium model. Results: We show that the blood kinetic energy predicted by the heterogeneous model is in strong agreement with the conventional approach. The homogeneity assumption, on the other hand, significantly over-predicts the blood kinetic energy within the aneurysmal sac. Conclusions: These results indicate that the benefits of the porous medium assumption can be retained if a heterogeneous approach is applied. Implementation of the presented method led to a substantial reduction in the total number of mesh elements compared to the conventional method, and greater accuracy was enabled by considering heterogeneity compared to the homogenous approach.
KW - Brain aneurysm
KW - CFD
KW - Embolic coils
KW - Heterogeneous
KW - Porous medium
UR - http://www.scopus.com/inward/record.url?scp=85061966773&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85061966773&partnerID=8YFLogxK
U2 - 10.1007/s13239-018-00383-1
DO - 10.1007/s13239-018-00383-1
M3 - Article
C2 - 30341728
AN - SCOPUS:85061966773
SN - 1869-408X
VL - 10
SP - 32
EP - 45
JO - Cardiovascular Engineering and Technology
JF - Cardiovascular Engineering and Technology
IS - 1
ER -